Fecal Occult Blood Testing: Laboratory Aspects of FIT
Pesquisa de sangue oculto nas fezes: aspectos laboratoriais do FIT
Mauren Isfer Anghebem1, Flávia Martinello2
1 Farmacêutica-Bioquímica; Doutorado em Ciências Farmacêuticas – Análises Clínicas; Professora Adjunta da Escola de Medicina e Ciências da Vida da Pontifícia Universidade Católica e Professora Adjunta do Departamento de Análises Clínicas da Universidade Federal do Paraná/UFPR. Curitiba, PR, Brazil; Membro do Grupo Técnico-Científico da Sociedade Brasileira de Análises Clínicas – SBAC.
2 Farmacêutica-Bioquímica; Pós-doutorado em Análises Clínicas; Professora do Departamento de Análises Clínicas da Universidade Federal de Santa Catarina/UFSC. Florianópolis, SC, Brazil; Membro do Grupo Técnico-Científico da Sociedade Brasileira de Análises Clínicas – SBAC.
Received on 08/26/2026
Approved on 08/26/2026
DOI: 10.21877/2448-3877.202600342.en
INTRODUCTION
Fecal occult blood testing aims to detect low-volume gastrointestinal bleeding that is not apparent on visual inspection and may be associated with a broad spectrum of clinical conditions, including hemorrhoids, erosive gastritis, esophagitis, peptic ulcers, angiodysplasias, and parasitic infections, as well as potentially more serious conditions such as inflammatory bowel diseases, adenomatous polyps, and colorectal neoplasms.(1-3)
In the context of colorectal cancer screening, fecal occult blood testing is recognized as a population-based screening tool rather than a definitive diagnostic test. A positive result indicates the presence of blood in the gastrointestinal tract but, by itself, is not pathognomonic for colorectal cancer because of its lack of specificity.(4) Following a positive result, colonoscopy is recommended as a confirmatory examination to determine the etiology of the bleeding, as recommended by current guidelines.(5,6)
Stool-based colorectal cancer screening tests include the guaiac-based fecal occult blood test (gFOBT), the fecal immunochemical test (FIT), and the multitarget stool DNA test (mt-sDNA). Direct visualization screening tests include computed tomography colonography, flexible sigmoidoscopy, and conventional colonoscopy. Another strategy combines flexible sigmoidoscopy with FIT, integrating a stool-based test with direct visualization. Emerging screening tests include colon capsule endoscopy (also referred to as capsule colonoscopy), blood-based screening tests (such as microRNAs and protein biomarkers), fecal microbiome studies, and urinary metabolomics.(6)
The traditional gFOBT is based on the chemical activity of the heme group and may react with hemoglobin from different sources. FIT, in turn, uses antibodies specific to the globin portion of human hemoglobin. This difference results in distinct characteristics regarding analytical specificity, interferences, and the number of specimens required. Compared with gFOBT, FIT has greater sensitivity for colorectal cancer and advanced neoplasia, with similar or slightly lower specificity, depending on the study. In addition, FIT is associated with higher screening adherence because it generally requires collection of a single specimen and does not impose dietary restrictions. Studies have demonstrated greater adherence and higher detection rates of advanced neoplasia in population-based programs using FIT.(7–9)
Consistent with organized colorectal cancer screening programs implemented in countries across Europe, the Western Pacific, East Asia, and the Americas, CONITEC (National Committee for Health Technology Incorporation into the Brazilian Unified Health System) recommends FIT as the preferred option for detecting occult blood in stool for colorectal cancer screening.(10) Accordingly, the Brazilian Ministry of Health included FIT for colorectal cancer screening in the SUS procedure table. This measure was published in the Brazilian Federal Official Gazette (Diário Oficial da União, in Portuguese) on August 10, 2026, and the new procedure will be registered in SUS information systems starting in September 2026.(11)
The present study is a narrative literature review aimed at synthesizing the main laboratory aspects related to fecal occult blood testing, with emphasis on FIT. The literature search was conducted in the PubMed/MEDLINE, SciELO, LILACS, and ScienceDirect databases. Controlled vocabulary terms (Occult Blood; Colorectal Neoplasms; Mass Screening; Immunologic Tests), selected from the Health Sciences Descriptors (DeCS) and Medical Subject Headings (MeSH), as well as free-text terms (Fecal Immunochemical Test [FIT]; Fecal Occult Blood Testing), were used in Portuguese and English. The inclusion criteria comprised original articles, literature reviews, clinical guidelines, and technical documents published between 1996 and 2026, available in full text, and addressing diagnostic, laboratory, and clinical aspects related to the topic. Conference abstracts and publications that did not present clinical or laboratory data relevant to the topic were excluded.
PRINCIPLES AND CHARACTERISTICS
OF THE METHODS
Guaiac-Based Method (gFOBT)
The gFOBT is based on the pseudoperoxidase activity of the heme group in hemoglobin. In the presence of hydrogen peroxide, heme catalyzes the oxidation of guaiac, producing a blue color. Thus, the test indirectly detects the presence of blood in stool through a chemical reaction and is not specific to human hemoglobin. However, the disadvantages of gFOBT include the need to collect specimens on three consecutive days, the inability to distinguish the source of bleeding (upper or lower gastrointestinal tract), and the need for dietary restrictions before testing. The diagnostic performance of gFOBT is inferior to that of the fecal immunochemical test (FIT), and many experts recommend replacing gFOBT with FIT. However, gFOBT has been shown to reduce colorectal cancer mortality and therefore represents a reasonable alternative when immunochemical tests are unavailable.(7,8,12)
Fecal Immunochemical Test (FIT)
FIT uses antibodies specific to the globin portion of human hemoglobin to detect and, in quantitative methods, measure the concentration of hemoglobin in stool. Hemoglobin originating from the upper gastrointestinal tract undergoes degradation during gastrointestinal transit, particularly in its globin portion, which is recognized by the antibodies, thereby reducing its detection by FIT. In contrast, hemoglobin originating from the colon and rectum tends to remain sufficiently intact to be recognized by the test, thereby increasing the specificity of FIT for colorectal bleeding. Unlike gFOBT, FIT does not depend on the pseudoperoxidase activity of heme and is not subject to the same interferences related to the consumption of meat or foods containing peroxidases. Specimen collection is performed using a dedicated collection device, generally with a single stool specimen per screening round.(7,8,13-15)
Colonoscopy has greater sensitivity than FIT and gFOBT; however, it is an invasive procedure and is associated with a risk of complications.(16)
Table 1 presents a comparison between the gFOBT (guaiac-based) and FIT (immunochemical) methods.
Table 1
Comparison between gFOBT and FIT.
| Characteristic | gFOBT | FIT |
| Analytical target | Chemical activity of heme | Human hemoglobin |
| Origin of detected blood | Human and animal | Human |
| Dietary restriction | Required | Not required |
| Medication-related interference | More frequent | Less frequent |
| Predominant source of bleeding | Gastrointestinal tract | Colon and rectum |
| Number of specimens | Usually three | One |
| Sensitivity for colorectal cancer | Lower | Higher |
Source: Adapted from 7,8,15.
COMPARISON OF FIT WITH STOOL DNA TESTING
The next-generation multitarget stool DNA test showed greater sensitivity than FIT for colorectal cancer and advanced precancerous lesions in the BLUE-C study, but lower specificity(17) A subsequently published meta-analysis showed a similar trend, with greater sensitivity for the multitarget stool DNA test and greater specificity for FIT.(18)
However, a Chinese multicenter study comparing two stool DNA tests with FIT found no significant advantage of the DNA tests over FIT when the tests were adjusted to equivalent specificity.(19)
QUALITATIVE AND QUANTITATIVE FIT
FIT can be performed using qualitative or quantitative methods. Qualitative FIT generally uses lateral flow immunochromatography, in which the presence of human hemoglobin above the device’s limit of detection produces a test line, allowing the specimen to be classified as reactive (positive) or nonreactive (negative). Visual reading, particularly when the intensity of the test line is close to the decision threshold, may be subject to variability in interpretation.(4,7)
Quantitative FIT uses automated analyzers, frequently employing immunoturbidimetric methods, to quantify the concentration of hemoglobin in stool, generally expressed as µg Hb/g of feces. By providing a quantitative measurement of fecal hemoglobin, FIT allows different cutoffs to be established for defining positivity. The cutoff may be adjusted according to test performance and the capacity of the screening program to perform colonoscopies. Automation also provides greater standardization of result interpretation, quality control, and high-throughput processing.(4,7)
Differences in performance exist between qualitative and quantitative FITs and should not be attributed exclusively to the test format. In a population-based study conducted in South Korea involving 1,181,904 participants, quantitative FIT had a lower positivity rate than qualitative FIT (2.5% vs. 8.1%) but was associated with a higher colorectal cancer detection rate among individuals who underwent colonoscopy, demonstrating that differences in positivity rates do not necessarily correspond to equivalent differences in the ability to detect neoplasia.(4,20,21) However, comparative studies have shown that diagnostic performance is also strongly influenced by the cutoff used. In a study comparing qualitative and quantitative FIT, when the cutoffs were adjusted to produce similar positivity rates, differences in performance between the methods were reduced, indicating that some of the observed differences may be related to the decision threshold rather than solely to the qualitative or quantitative test format.(22)
Thus, the choice between qualitative and quantitative FIT should consider not only the analytical performance of the device but also the cutoff used, the screening strategy, available resources, processing capacity, and, most importantly, the availability of colonoscopy for the investigation of positive results.(4,7,21)
CUTOFF AND LABORATORY IMPLICATIONS
The cutoff is a key variable affecting FIT performance and the planning of screening programs. In quantitative tests, the positivity threshold may be adjusted according to program objectives and the available capacity for colonoscopic investigation. In general, increasing the cutoff reduces the proportion of positive results and, consequently, the demand for colonoscopies, but is associated with reduced sensitivity for detecting colorectal cancer and precursor lesions.(23,24) A systematic review and meta-analysis involving 46 studies and approximately 2.4 million participants demonstrated that increasing the FIT cutoff is associated with reduced sensitivity for colorectal cancer and advanced adenomas, accompanied by increased specificity.(23,24)
FIT cutoffs used internationally vary widely. In a survey of 38 population-based screening programs using quantitative FIT, thresholds ranged from 8.5 to 120 µg Hb/g of feces, with 20 µg Hb/g being the most frequently used value. Threshold selection was primarily related to the need to maximize sensitivity for colorectal cancer and precursor lesions or, alternatively, to align demand with the available capacity to perform colonoscopies.(24) A subsequent international study demonstrated that this wide variation in cutoffs is associated with important differences in positivity rates and the diagnostic performance of screening programs: at the extremes evaluated, sensitivity for colorectal cancer ranged from 98.1% to 55.6%, whereas specificity ranged from 86.3% to 98.9%.(23,24)
In Brazil, the Preliminary Report on the Brazilian Guidelines for Colorectal Cancer Screening issued by CONITEC recommends the use of qualitative or quantitative FIT for asymptomatic, average-risk individuals aged 50 to 75 years, with a cutoff of 50 ng/mL. The document states that this value corresponds to approximately 10 µg Hb/g of feces. This standardization allows comparison with values used in international programs, provided that the specific characteristics of the collection device and analytical system are taken into account.(10)
The unit used to express the FIT cutoff depends on the collection system and analytical method. A concentration expressed in ng/mL represents the amount of hemoglobin in the extraction buffer, whereas µg Hb/g of feces estimates the concentration of hemoglobin in fecal material, taking into account the amount of stool transferred and the volume of buffer used. Therefore, conversion between these units should not be performed by simple dimensional equivalence; the relationship established for the collection device and analytical system used should be applied.(10)
The US Multi-Society Task Force on Colorectal Cancer, comprising experts from three US gastroenterology organizations, namely, the American Gastroenterological Association, the American College of Gastroenterology, and the American Society for Gastrointestinal Endoscopy, recommends that the cutoff be defined according to the performance characteristics of the test, the device used, cost-effectiveness, and the available capacity to perform colonoscopies. The Task Force suggests a threshold of ≤20 µg Hb/g of feces, although this is a weak recommendation based on low-quality evidence.(15)
Regarding screening intervals, CONITEC recommends FIT every two years for asymptomatic, average-risk individuals aged 50 to 75 years. This is a conditional recommendation based on low-certainty evidence and on acceptability and feasibility criteria. The report emphasizes that no randomized clinical trials or systematic reviews directly comparing annual and biennial FIT were identified; therefore, the biennial recommendation also takes into account indirect evidence, experience from international programs, and considerations related to the feasibility and sustainability of screening.(10)
The heterogeneity of cutoffs across screening programs underscores that FIT results should not be interpreted independently of the analytical system, collection device, reporting unit, cutoff, and screening strategy. Thus, fecal hemoglobin concentration should always be interpreted in the context of the method and cutoff used by the screening program or specified by the manufacturer.
VARIABILITY AMONG METHODS AND BRANDS
There is no harmonization among commercially available FIT assays. Performance may vary across brands, even when nominally similar cutoffs are used. Differences in antibodies, buffers, and collection devices contribute to this variability.(25,26)
A study comparing five FITs demonstrated an approximately fourfold variation in positivity rates among brands, as well as significant differences in sensitivity and specificity. The authors concluded that different FITs should not be regarded as interchangeable methods.(26)
A meta-analysis of different brands of quantitative FIT showed, however, that some of the heterogeneity observed among assays may be related to the cutoff used. The positivity rate showed a strong correlation with sensitivity and an inverse relationship with specificity.(27) Thus, both the assay-specific analytical characteristics and the definition of the cutoff should be considered when comparing methods.
PREANALYTICAL ASPECTS
Fecal Hemoglobin Stability
The stability of fecal hemoglobin represents an important preanalytical limitation of FIT. Because the method uses antibodies directed against the globin portion of hemoglobin, the integrity of this component is essential for reliable analyte detection. Globin is relatively unstable and undergoes progressive degradation after specimen collection, which may compromise hemoglobin recovery and, consequently, the test result.(28) In specimens with hemoglobin concentrations close to the cutoff, this degradation may have a greater impact on the result. A progressive decrease in positivity has been observed in specimens stored at room temperature, whereas refrigeration helped preserve hemoglobin stability, although it did not completely prevent hemoglobin loss over time.(28)
Stability may also vary among different commercial systems. A study comparing different kits showed that hemoglobin recovery was influenced by storage time and temperature. Likewise, freeze–thaw cycles produced heterogeneous effects across systems, indicating that storage conditions should not be generalized to all devices.(29)
Studies conducted in population-based screening programs have indicated that specimen storage for periods of up to 10 days did not always result in significant changes in positivity rates. These findings do not diminish the importance of controlling storage time and conditions, particularly because hemoglobin degradation may reduce the detectable hemoglobin concentration and primarily affect specimens with values close to the cutoff. Therefore, specimens should be transported and processed as promptly as possible, in accordance with the storage conditions established for the system used.(4)
Specimen Collection
The amount of stool introduced into the collection device is an important preanalytical factor in FIT because it affects the relationship between the amount of stool sampled and the volume of buffer and, consequently, the measured hemoglobin concentration. Undersampling may reduce hemoglobin recovery and increase the risk of false-negative results, whereas an excessive amount of stool may alter the measured concentration and interfere with the result. Specimen collection should therefore strictly follow the technique recommended by the manufacturer, with standardized patient instructions regarding the appropriate specimen amount and the correct use of the collection device.(30,31)
Medication-Related Interference
FIT is less affected by dietary and medication-related factors than gFOBT.(7) However, evidence regarding the use of acetylsalicylic acid (ASA), nonsteroidal anti-inflammatory drugs (NSAIDs), antiplatelet agents, and anticoagulants is heterogeneous. While one meta-analysis found no significant difference in positive predictive value between users and nonusers of these medications, other studies reported a reduced positive predictive value for advanced colorectal neoplasia, particularly among users of ASA, antiplatelet agents, and oral anticoagulants.(7,32) In a larger study, users of ASA and direct oral anticoagulants also had a lower positive predictive value.(33)
Despite these associations, routine discontinuation of antithrombotic therapy before FIT is not recommended. The use of these medications should be considered when interpreting the results but, by itself, does not constitute a reason to disregard a positive result, which should be followed by the indicated diagnostic evaluation.(24,33)
ANALYTICAL ASPECTS
In qualitative FITs based on lateral flow immunochromatography, very high hemoglobin concentrations may produce the hook effect, characterized by a paradoxical reduction in test-line intensity and the possibility of a false-negative result. This phenomenon represents a limitation of qualitative systems and reinforces the advantages of quantitative methods when available.(34)
Another limitation is related to hemoglobin variants. Different FITs may vary in their ability to recognize these variants, and some variants were detected less efficiently by all systems evaluated, potentially leading to false-negative results in certain subgroups.(14) In addition, differences in the ability of antibodies to recognize partially degraded forms of globin may contribute to differences in sensitivity among platforms, even when the same cutoff is used.(35)
These findings reinforce that quantitative results obtained using different platforms should not automatically be considered interchangeable and should be interpreted according to the analytical system and cutoff used.(35)
QUALITY CONTROL
Quality control of FIT presents challenges related to the lack of standardization among methods, the absence of an international reference material for fecal hemoglobin, and the difficulty of developing control materials that adequately reproduce the characteristics of stool.(25)
Internal Quality Control
The availability of manufacturer-independent control materials for internal quality control is limited. This limitation hampers the daily assessment of analytical performance. The Fecal Immunochemical Test Working Group of the International Federation of Clinical Chemistry and Laboratory Medicine has proposed alternatives for the development of third-party control materials, but their availability remains limited.(25,28)
Comparative studies of automated quantitative systems have demonstrated good precision, recovery, and linearity, but have also identified differences in control results among platforms.(36)
External Quality Assessment
An international survey identified 24 external quality assessment schemes for FIT, with responses obtained for 16. Differences were observed in the matrix used, the buffer employed, and the units used to report results.(37) Stool-like matrices showed greater variability, primarily because of material heterogeneity, whereas liquid matrices, although more stable, are less representative of clinical specimen conditions.(38)
As an alternative, the use of artificial stool containing stable hemoglobin and an internal standard has been proposed, allowing estimation of specimen weight and correction of the measured hemoglobin concentration.(39)
In the absence of fully standardized materials and procedures across platforms, external quality assessment should primarily be regarded as a tool for monitoring the stability and consistency of each system’s performance over time, rather than as a means of directly comparing results across different methods.(25)
METHOD VERIFICATION AND VALIDATION
Verification or validation of a FIT method should follow the general principles applicable to quantitative methods in clinical chemistry, while taking into account the specific characteristics of the assay.(25,40) Relevant parameters include the limit of detection, limit of quantification, precision, reproducibility, preanalytical variables, and method comparison.(41)
The cutoff reported by the manufacturer should be evaluated within the context of the system used, taking into account the characteristics of the antibodies, buffers, and collection devices. In the absence of a primary reference material and standardization among manufacturers, equivalence among different systems should not be assumed.(25,40)
The evaluation should also include the reproducibility of sampling and measurement and, when applicable, agreement between results obtained by automated and visual reading.(25)
When replacing one kit with another, the comparability of the results with those obtained using the previously employed method should be evaluated, taking into account the characteristics of each system and, when relevant, their relationship with colonoscopic findings.(26,42)
INTERPRETATION OF RESULTS
Hemorrhoids
The contribution of hemorrhoids to false-positive FIT results remains uncertain. Studies have reported different findings: while a Korean study identified a significant association between hemorrhoids and false-positive results,(43) a Dutch study found a much smaller contribution, with isolated hemorrhoids present in only a small proportion of these cases.(44) Another study identified several factors associated with a positive FIT in the absence of neoplasia, without demonstrating a strong association with hemorrhoids alone.(45)
Among individuals with rectal bleeding, FIT also shows a high ability to detect colorectal cancer. A Swedish study demonstrated similar test performance in individuals with and without a history of rectal bleeding.(46) In the NICE FIT study, undetectable fecal hemoglobin concentrations were associated with a low risk of colorectal cancer in patients with rectal bleeding,(47) a finding consistent with a meta-analysis demonstrating high FIT sensitivity in symptomatic patients.(48)
Therefore, neither the presence of hemorrhoids nor rectal bleeding should be used in isolation to interpret a positive or negative FIT result; the result should be considered within the clinical context and, when positive, should be followed by the indicated diagnostic evaluation.
Inflammatory Bowel Disease
FIT is associated with endoscopic disease activity in inflammatory bowel disease, particularly in ulcerative colitis, but its performance is lower in Crohn’s disease with isolated ileal involvement.(41)
In a comparative study, fecal calprotectin showed greater sensitivity for endoscopic and histological activity, whereas FIT showed greater specificity. In isolated ileal Crohn’s disease, FIT sensitivity was particularly low compared with that of fecal calprotectin, whereas its performance was better in cases with colonic involvement.(41)
In ulcerative colitis, studies have demonstrated a correlation between fecal hemoglobin concentration and endoscopic activity, including an association with mucosal healing in certain settings.(49,50)
Colorectal Cancer
A Cochrane systematic review demonstrated superior performance of FIT compared with gFOBT in discriminating advanced colorectal neoplasia and colorectal cancer. At prespecified specificities of 90% and 95%, FIT sensitivities were 0.97 and 0.89, respectively, compared with 0.80 and 0.73 for gFOBT.(7) Randomized clinical trials of gFOBT have also demonstrated reductions in colorectal cancer mortality and incidence, although equivalent randomized head-to-head trials evaluating FIT are not available.(8)
Interpretation of FIT should take into account the characteristics of the method, the cutoff used, and preanalytical conditions. A positive result indicates the presence of fecal hemoglobin but does not confirm colorectal cancer; hemorrhoids, inflammation, and other conditions may also be associated with FIT positivity.(43-48)
In patients taking anticoagulants or antiplatelet agents, these medications should not constitute a reason to disregard a positive FIT result because, although they may reduce the positive predictive value in some populations, the association between FIT positivity and colorectal neoplasia remains relevant.(33,51)
Fecal hemoglobin concentration may also vary according to population characteristics, such as sex and age, as well as geographic region and the FIT system used.(52) Elevated values may occur in the absence of an identifiable colorectal lesion and in different conditions associated with inflammation.(52)
Therefore, laboratory results should be interpreted within the clinical context and in conjunction with the specific characteristics of the method and analytical system used.
FINAL CONSIDERATIONS
FIT is a sensitive and specific immunochemical method for detecting human hemoglobin in stool and offers advantages over gFOBT, including the absence of dietary restrictions and the possibility of collecting a single specimen.(7,8,15)
The lack of harmonization among commercial systems represents an important limitation. Different antibodies, devices, and cutoffs may produce different results; therefore, different FITs should not automatically be considered interchangeable. In addition, clinical and biological factors may also influence result interpretation.(14,25-27,32,33,41,43,44,46-48,51,52)
Thus, the quality of fecal occult blood testing depends not only on hemoglobin detection but also on the integration of a controlled preanalytical phase, analytical performance, quality control, an appropriate cutoff, and context-based interpretation of results.
REFERENCES
- Jacobsson M, Wagner V, Kanneganti S. Screening for Colorectal Cancer. Surg Clin North Am. 2024;104(3):595-607. doi:10.1016/j.suc.2023.11.009
- Lee JK, Liles EG, Bent S, Levin TR, Corley DA. Accuracy of fecal immunochemical tests for colorectal cancer: systematic review and meta-analysis. Ann Intern Med. 2014;160(3):171. doi:10.7326/M13-1484
- Maryam Faridi, Pragyan Swagatika Panda, Man Mohan Mehndiratta, Megha Rastogi, Swati Sharma, Nazia Khan. Fecal occult blood test and its correlation with Gastrointestinal parasites. International Journal of Scholarly Research in Multidisciplinary Studies. 2023;2(2):008-012. doi:10.56781/ijsrms.2023.2.2.0054
- Robertson DJ, Lee JK, Boland CR, et al. Recommendations on Fecal Immunochemical Testing to Screen for Colorectal Neoplasia: A Consensus Statement by the US Multi-Society Task Force on Colorectal Cancer. Gastroenterology. 2017;152(5):1217-1237.e3. doi:10.1053/j.gastro.2016.08.053
- Chen Y, Zhang Y, Yan Y, et al. Global colorectal cancer screening programs and coverage rate estimation: an evidence synthesis. J Transl Med. 2025;23(1):811. doi:10.1186/s12967-025-06887-4
- Jain S, Maque J, Galoosian A, Osuna-Garcia A, May FP. Optimal Strategies for Colorectal Cancer Screening. Curr Treat Options Oncol. 2022;23(4):474-493. doi:10.1007/s11864-022-00962-4
- Grobbee EJ, Wisse PH, Schreuders EH, et al. Guaiac-based faecal occult blood tests versus faecal immunochemical tests for colorectal cancer screening in average-risk individuals. Cochrane Database Syst Rev. 2022;6(6):CD009276. doi:10.1002/14651858.CD009276.pub2
- Lieberman D, Ladabaum U, Cruz-Correa M, et al. Screening for Colorectal Cancer and Evolving Issues for Physicians and Patients: A Review. JAMA. 2016;316(20):2135-2145. doi:10.1001/jama.2016.17418
- Burke CA, Lieberman D, Feuerstein JD. AGA Clinical Practice Update on Approach to the Use of Noninvasive Colorectal Cancer Screening Options: Commentary. Gastroenterology. 2022;162(3):952-956. doi:10.1053/j.gastro.2021.09.075
- Brasil. Ministério da Saúde. Relatório preliminar – Diretrizes Brasileiras do Rastreamento do câncer de cólon e reto. Comissão Nacional de Incorporação de Tecnologias no Sistema Único de Saúde (CONITEC). 2026.. Available from: https://www.gov.br/conitec/pt-br/midias/consultas/relatorios/2026/relatorio-preliminar-diretrizes-brasileiras-do-rastreamento-do-cancer-de-colon-e-reto-cp-20. Accessed: August 18, 2026.
- Instituto Nacional de Câncer (INCA). SUS inclui teste de alta tecnologia para rastreamento de câncer colorretal. Available from: https://www.gov.br/inca/pt-br/assuntos/noticias/2026/sus-inclui-teste-de-alta-tecnologia-para-rastreamento-de-cancer-colorretal. Accessed: August 10, 2026.
- Allison JE, Tekawa IS, Ransom LJ, Adrain AL. A comparison of fecal occult-blood tests for colorectal-cancer screening. N Engl J Med. 1996;334(3):155-159. doi:10.1056/NEJM199601183340304
- Robertson DJ, Selby K. Fecal Immunochemical Test: The World’s Colorectal Cancer Screening Test. Gastrointest Endosc Clin N Am. 2020;30(3):511-526. doi:10.1016/j.giec.2020.02.011
- Carroll MR, John C, Mantio D, Djedovic NK, Benton SC. An assessment of the effect of haemoglobin variants on detection by faecal immunochemical tests. Ann Clin Biochem. 2018;55(6):706-709. doi:10.1177/0004563218778716
- Burke CA, Lieberman D, Feuerstein JD. AGA Clinical Practice Update on Approach to the Use of Noninvasive Colorectal Cancer Screening Options: Commentary. Gastroenterology. 2022;162(3):952-956. doi:10.1053/j.gastro.2021.09.075
- Knudsen AB, Zauber AG, Rutter CM, et al. Estimation of Benefits, Burden, and Harms of Colorectal Cancer Screening Strategies: Modeling Study for the US Preventive Services Task Force. JAMA. 2016;315(23):2595-2609. doi:10.1001/jama.2016.6828
- Imperiale TF, Porter K, Zella J, et al. Next-Generation Multitarget Stool DNA Test for Colorectal Cancer Screening. N Engl J Med. 2024;390(11):984-993. doi:10.1056/NEJMoa2310336
- Ebner DW, Johnson HA, Estes C, et al. Multi-Target Stool DNA and the Fecal Immunochemical Test: A Systematic Review and Meta-Analysis on Test Performances. Am J Prev Med. 2025;69(2):107654. doi:10.1016/j.amepre.2025.107654
- Jin P, You P, Fang J, et al. Comparison of Performance of Two Stool DNA Tests and a Fecal Immunochemical Test in Detecting Colorectal Neoplasm: A Multicenter Diagnostic Study. Cancer Epidemiol Biomarkers Prev. 2022;31(3):654-661. doi:10.1158/1055-9965.EPI-21-0991
- Selby K, Levine EH, Doan C, et al. Effect of Sex, Age, and Positivity Threshold on Fecal Immunochemical Test Accuracy: A Systematic Review and Meta-analysis. Gastroenterology. 2019;157(6):1494-1505. doi:10.1053/j.gastro.2019.08.023
- Park MJ, Choi KS, Lee YK, Jun JK, Lee HY. A comparison of qualitative and quantitative fecal immunochemical tests in the Korean national colorectal cancer screening program. Scand J Gastroenterol. 2012;47(4):461-466. doi:10.3109/00365521.2012.668930
- Lu M, Zhang YH, Lu B, et al. Head-to-head comparison of the test performance of self-administered qualitative vs. laboratory-based quantitative fecal immunochemical tests in detecting colorectal neoplasm. Chin Med J (Engl). 2021;134(11):1335-1344. doi:10.1097/CM9.0000000000001524
- Ola I, Seum T, Carlsson S V, Hoffmeister M, Brenner H. Variation in diagnostic performance of fecal immunochemical test (FIT) by cutoffs used in screening programs globally. EClinicalMedicine. 2026;94:103835. doi:10.1016/j.eclinm.2026.103835
- Kaminski MF, Robertson DJ, Senore C, Rex DK. Optimizing the Quality of Colorectal Cancer Screening Worldwide. Gastroenterology. 2020;158(2):404-417. doi:10.1053/j.gastro.2019.11.026
- Benton SC, Symonds E, Djedovic N, et al. Faecal immunochemical tests for haemoglobin: Analytical challenges and potential solutions. Clin Chim Acta. 2021;517:60-65. doi:10.1016/j.cca.2021.01.024
- Levy BT, Xu Y, Daly JM, et al. Comparative Performance of Common Fecal Immunochemical Tests : A Cross-Sectional Study. Ann Intern Med. 2024;177(10):1350-1360. doi:10.7326/M24-0080
- Gies A, Bhardwaj M, Stock C, Schrotz-King P, Brenner H. Quantitative fecal immunochemical tests for colorectal cancer screening. Int J Cancer. 2018;143(2):234-244. doi:10.1002/ijc.31233
- Reid MS, Paul HA, Mostoufi A, Robinson JL, Sadrzadeh SMH. Evaluation of the stability of fecal immunochemical test specimens. Clin Biochem. 2023;115:92-96. doi:10.1016/j.clinbiochem.2022.11.015
- Catomeris P, Baxter NN, Boss SC, et al. Effect of Temperature and Time on Fecal Hemoglobin Stability in 5 Fecal Immunochemical Test Methods and One Guaiac Method. Arch Pathol Lab Med. 2018;142(1):75-82. doi:10.5858/arpa.2016-0294-OA
- Gies A, Gruner LF, Schrotz-King P, Brenner H. Effect of Imperfect Compliance With Instructions for Fecal Sample Collection on Diagnostic Performance of 9 Fecal Immunochemical Tests. Clin Gastroenterol Hepatol. 2019;17(9):1829-1839.e4. doi:10.1016/j.cgh.2019.03.001
- Symonds EL, Fraser CG, Bastin D, Berwald G, Young GP. The Effect of the Variability in Fecal Immunochemical Test Sample Collection Technique on Clinical Performance. Cancer Epidemiol Biomarkers Prev. 2021;30(1):175-181. doi:10.1158/1055-9965.EPI-20-0984
- Jung YS, Im E, Park CH. Impact of antiplatelet agents and anticoagulants on the performance of fecal immunochemical tests: a systematic review and meta-analysis. Surg Endosc. 2022;36(6):4299-4311. doi:10.1007/s00464-021-08774-7
- Randel KR, Botteri E, Romstad KMK, et al. Effects of Oral Anticoagulants and Aspirin on Performance of Fecal Immunochemical Tests in Colorectal Cancer Screening. Gastroenterology. 2019;156(6):1642-1649.e1. doi:10.1053/j.gastro.2019.01.040
- Ross GMS, Filippini D, Nielen MWF, Salentijn GI. Unraveling the Hook Effect: A Comprehensive Study of High Antigen Concentration Effects in Sandwich Lateral Flow Immunoassays. Anal Chem. 2020;92(23):15587-15595. doi:10.1021/acs.analchem.0c03740
- Chiang TH, Chuang SL, Chen SLS, et al. Difference in performance of fecal immunochemical tests with the same hemoglobin cutoff concentration in a nationwide colorectal cancer screening program. Gastroenterology. 2014;147(6):1317-1326. doi:10.1053/j.gastro.2014.08.043
- Piggott C, Carroll MRR, John C, O’Driscoll S, Benton SC. Analytical evaluation of four faecal immunochemistry tests for haemoglobin. Clin Chem Lab Med. 2020;59(1):173-178. doi:10.1515/cclm-2020-0251
- Piggott C, Thelen M, Jones S, et al. A survey of external quality assessment schemes (EQAS) for faecal immunochemical tests for haemoglobin (FIT). Clin Chim Acta. 2026;592:121180. doi:10.1016/j.cca.2026.121180
- O’Driscoll S, Piggott C, Bruce H, Benton SC. An evaluation of ten external quality assurance scheme (EQAS) materials for the faecal immunochemical test (FIT) for haemoglobin. Clin Chem Lab Med. 2020;59(2):307-313. doi:10.1515/cclm-2020-0210
- Yasui R, Yamada M, Takehara S, Sakurabayashi I, Watanabe K. Novel artificial stool material for external quality assurance (EQA) on a fecal immunochemical test for hemoglobin (FIT): The confirmed utility of stable hemoglobin and an internal standard material. Clin Chim Acta. 2018;483:76-81. doi:10.1016/j.cca.2018.04.021
- Godber IM, Benton SC, Fraser CG. Setting up a service for a faecal immunochemical test for haemoglobin (FIT): a review of considerations, challenges and constraints. J Clin Pathol. 2018;71(12):1041-1045. doi:10.1136/jclinpath-2018-205047
- Bohra A, Batt N, Dutt K, et al. The synergy of dual faecal immunochemical and faecal calprotectin testing for accurate assessment of endoscopic and histological activity in inflammatory bowel disease. Therap Adv Gastroenterol. 2024;17:17562848241237896. doi:10.1177/17562848241237895
- Wieten E, de Klerk CM, van der Steen A, et al. Equivalent Accuracy of 2 Quantitative Fecal Immunochemical Tests in Detecting Advanced Neoplasia in an Organized Colorectal Cancer Screening Program. Gastroenterology. 2018;155(5):1392-1399.e5. doi:10.1053/j.gastro.2018.07.021
- Kim NH, Park JH, Park D Il, Sohn C Il, Choi K, Jung YS. Are Hemorrhoids Associated with False-Positive Fecal Immunochemical Test Results? Yonsei Med J. 2017;58(1):150-157. doi:10.3349/ymj.2017.58.1.150
- van Turenhout ST, Oort FA, Terhaar sive Droste JS, et al. Hemorrhoids detected at colonoscopy: an infrequent cause of false-positive fecal immunochemical test results. Gastrointest Endosc. 2012;76(1):136-143. doi:10.1016/j.gie.2012.03.169
- Amitay EL, Cuk K, Niedermaier T, Weigl K, Brenner H. Factors associated with false-positive fecal immunochemical tests in a large German colorectal cancer screening study. Int J Cancer. 2019;144(10):2419-2427. doi:10.1002/ijc.31972
- Högberg C, Gunnarsson U, Cronberg O, Thulesius H, Lilja M, Jansson S. Qualitative faecal immunochemical tests (FITs) for diagnosing colorectal cancer in patients with histories of rectal bleeding in primary care: a cohort study. Int J Colorectal Dis. 2020;35(11):2035-2040. doi:10.1007/s00384-020-03672-1
- Hicks G, D’Souza N, Georgiou Delisle T, et al. Using the faecal immunochemical test in patients with rectal bleeding: evidence from the NICE FIT study. Colorectal Dis. 2021;23(7):1630-1638. doi:10.1111/codi.15593
- Saw KS, Liu C, Xu W, Varghese C, Parry S, Bissett I. Faecal immunochemical test to triage patients with possible colorectal cancer symptoms: meta-analysis. Br J Surg. 2022;109(2):182-190. doi:10.1093/bjs/znab411
- Ryu DG, Kim HW, Park SB, et al. Clinical implications of fecal calprotectin and fecal immunochemical test on mucosal status in patients with ulcerative colitis. Medicine. 2019;98(36):e17080. doi:10.1097/MD.0000000000017080
- Takashima S, Kato J, Hiraoka S, et al. Evaluation of Mucosal Healing in Ulcerative Colitis by Fecal Calprotectin Vs. Fecal Immunochemical Test. Am J Gastroenterol. 2015;110(6):873-880. doi:10.1038/ajg.2015.66
- Nieuwenburg SA V, Vuik FER, Kruip MJHA, Kuipers EJ, Spaander MCW. Effect of anticoagulants and NSAIDs on accuracy of faecal immunochemical tests (FITs) in colorectal cancer screening: a systematic review and meta-analysis. Gut. 2019;68(5):866-872. doi:10.1136/gutjnl-2018-316344
- Fraser CG. Faecal haemoglobin: Measurement, applications, and future potential. Best Pract Res Clin Gastroenterol. 2023;66:101833. doi:10.1016/j.bpg.2023.101833